The RAM GENE-1 mark II, designed for work in extreme conditions, features a ruggedized casing and backlight.
This versatile instrument is a combined dose rate and contamination meter designed to be used with one hand.
NEW! Now available with external power plug which can also be used to charge compatible batteries
The RAM GENE-1 is equipped with a recessed, internal GM pancake detector using a thin mica window protected by a stainless steel screen.
The transition between measuring dose rate radiation and contamination is achieved by simply removing or placing the plastic cap.
Click here to see an online example of calibrating the RAM GENE-1 mark II
We use two Test Sources: Sr-90/Y-90 and Cs-137 190 mCi
Example of Calibration Certificate using two calibration points:
Beta (Cap off) at 64.7 Kcpm and gamma (Cap on) at 1080 uSv/h
Values can vary around 20% each direction
Click Here to see a comparison between the RAM GENE-1 mark II and Inspector
Detector: 1.75″ Pancake GM tube – LND73118 or equivalent
Count Range: 0 – 42,000 cps
Measuring Range: 0.05 µSv/hr to 7 mSv/hr (5 µR/hr to 700 mR/hr)
Sensitivity: Approximately 5.8 cps/µSv/hr (58 cps/mR/hr) – 137Cs